• Users Online: 616
  • Print this page
  • Email this page
ORIGINAL ARTICLE
Year : 2021  |  Volume : 29  |  Issue : 4  |  Page : 264-269

Automated follicular assessment using a novel two-dimensional ultrasound-based solution


1 Philips Research India, Philips Innovation Campus, Manyata Tech-Park, Bengaluru, Karnataka, India
2 Philips Research India, Philips Innovation Campus, Manyata Tech-Park, Bengaluru, Karnataka; KLA Corporation, Chennai, Tamil Nadu, India
3 Gunasheela Surgical and Maternity Hospital, Bengaluru, Karnataka, India

Correspondence Address:
Dr. Celine Firtion
Philips Research India, Philips Innovation Campus, Manyata Tech-Park, Bengaluru - 560 045, Karnataka
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/JMU.JMU_149_20

Rights and Permissions

Background: High intra- and interobserver variability in the follicular assessment using two-dimensional (2D) ultrasound (US) is still a concern. To solve this issue, we have developed a novel software solution, which automatically provides follicles' count and their diameters using 2D US images obtained by a manual sweep of an ovary. The primary objective of this study was to compare the result of the automated solution with a manual 2D US-based assessment. Methods: In the first phase, multiple follicular US sweeps were collected from 54 subjects; these sweeps were used to develop the software. In the second phase, data from 10 subjects were collected for validation of the developed solution. During each phase, for follicles ≥5 mm, their count and diameters were recorded by the sonologist using 2D US. Results: For the total follicle count, a high correlation (0.787) was observed between the solution and manual assessment. The 95% limits of agreement between the two methods were in the range of 4.232 to −4.258. The two methods had an excellent correlation (0.817) for the measurement of mean follicular diameter. However, the solution had a tendency to underestimate the mean diameter by an average of 1.725 mm (±2.16 mm). The limits of agreement between the two methods for mean diameter measurement were from 2.508 to −5.960 mm. Conclusion: This study validates the feasibility of our solution for automatic assessment of follicle count and diameter with accuracy comparable to the 2D US-based manual assessment. We further observed that the solution's performance is better than known intra- and interobserver variability of the manual assessment. We recommend further validation of the solution to confirm these initial results and potential time gain with an automated assessment.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed1833    
    Printed52    
    Emailed0    
    PDF Downloaded131    
    Comments [Add]    

Recommend this journal