ORIGINAL ARTICLE
Year : 2018  |  Volume : 26  |  Issue : 1  |  Page : 24-30

High-intensity focused ultrasound thermal lesion detection using entropy imaging of ultrasound radio frequency signal time series


1 Department of Biomedical Engineering, School of Electrical Engineering, Iran University of Science and Technology, Tehran, Iran
2 Department of Biomedical Engineering, Science and Research Branch, Islamic Azad University, Tehran, Iran
3 Department of Physics, Ryerson University; Keenan Research Centre for Biomedical Science, Institute for Biomedical Engineering, Science and Technology, St. Michael's Hospital, Toronto, ON, Canada

Date of Submission06-Feb-2017
Date of Acceptance05-May-2017
Date of Web Publication28-Mar-2018

Correspondence Address:
Dr. Hamid Behnam
Department of Biomedical Engineering, School of Electrical Engineering, Iran University of Science and Technology, Tehran 16844
Iran
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/JMU.JMU_3_17

Rights and Permissions
  Abstract 

Background: During the past few decades, high-intensity focused ultrasound (HIFU) modality has been gaining surging interest in various therapeutic applications such as non- or minimally-invasive cancer treatment. Among other attributes, robust and real-time HIFU treatment monitoring and lesion detection have become essential issues for successful clinical acceptance of the modality. More recently, ultrasound radio frequency (RF) time series imaging has been studied by a number of researchers. Materials and Methods: The objective of this study is to investigate the applicability of entropy parameter of RF time series of ultrasound backscattered signals, a. k. a. Entropy imaging, toward HIFU thermal lesion detection. To this end, five fresh ex vivo porcine muscle tissue samples were exposed to HIFU exposures with total acoustic powers ranging from 30 to 110 Watts. The contrast-to-speckle ratio (CSR) values of the entropy images and their corresponding B-mode images of pre-, during- and post-HIFU exposure for each acoustic power were calculated. Results: The novelty of this study is the use of Entropy parameter on ultrasound RF time series for the first time. Statistically significant differences were obtained between the CSR values for the B mode and entropy images at various acoustic powers. In case of 110 Watt, a CSR value 3.4 times higher than B-mode images was accomplished using the proposed method. Furthermore, the proposed method is compared with the scaling parameter of Nakagami imaging and same data which are used in this study. Conclusion: Entropy has the potential for using as an imaging parameter for differentiating lesions in HIFU surgery.

Keywords: Contrast-to-speckle ratio, entropy imaging, high intensity focused ultrasound, ultrasound radio frequency time series analysis


How to cite this article:
Monfared MM, Behnam H, Rangraz P, Tavakkoli J. High-intensity focused ultrasound thermal lesion detection using entropy imaging of ultrasound radio frequency signal time series. J Med Ultrasound 2018;26:24-30

How to cite this URL:
Monfared MM, Behnam H, Rangraz P, Tavakkoli J. High-intensity focused ultrasound thermal lesion detection using entropy imaging of ultrasound radio frequency signal time series. J Med Ultrasound [serial online] 2018 [cited 2023 Apr 2];26:24-30. Available from: http://www.jmuonline.org/text.asp?2018/26/1/24/228855


  Introduction Top


In recent years, a number of technological innovations have come to practice to reduce the level of invasiveness in various surgical procedures. High-intensity focused ultrasound (HIFU) is a notable example of such techniques applied in non- or–minimally-invasive treatments in areas such as oncology, general surgery, and cosmetic surgery.[1],[2],[3]

In a typical HIFU treatment, a thermal lesion can be produced in a region of tissue where a tumor exists. It results in an irreversible tissue coagulation necrosis in the region of interest (ROI). Thus, monitoring the process of creating a thermal lesion is very important for controlling the position of the probe, coagulated volume, lesion size, and also estimating the degree of coagulation. Some of the methods for monitoring and controlling the HIFU lesion creation process in tissue are using X-ray, computed tomography, and magnetic resonance imaging.[4] The latter provides high quality and high spatial resolution images. However, imaging by these methods is time-consuming and relatively expensive in comparison with ultrasound imaging. On the other hand, the real-time feedback for highly accurate positioning, heating control, high temporal resolution, low cost, and safety of the ultrasound imaging method have made it common and useful for HIFU treatment monitoring and control.[5]

Generally, lesion detection during a typical HIFU thermal therapy is more feasible than after HIFU exposure. Creation and collapsing of microscopic bubbles in the focal region during thermal induction of HIFU is the cavitation effect in the tissue that provides an effective method [6] of HIFU focal region detection.

In the past few decades, ultrasound tissue characterization, in general, and HIFU lesion detection, in particular, have been active topics of research. The basis of most ultrasound-based tissue characterization methods is that the backscattered ultrasound radio frequency (RF) echoes carry tissue dependent information that can be used for detecting the type of tissue.

A few studies have been conducted for analyzing ultrasound RF data that made the basis of the method proposed in our study. Moradi et al.[7] separated different types of animal tissues using RF echo signals acquired from a tissue sample. Imani et al.[8] proved that ultrasound RF time series imaging could correctly distinguish the ablated from nonablated tissue.

In two articles authored by Hughes,[9],[10] he verified that ultrasound signals could be analyzed with the use of entropy to characterize the changes in the microstructures of scattering media such as biological tissue. If the HIFU thermal lesion is formed, the structures of backscattered RF echoes in the same point at pre- and post-HIFU exposures are become significantly different. The main purposes of this study were (i) examining the entropy parameter of the backscattered RF echoes to estimate the location of the lesion, and (ii) comparing the contrast-to-speckle ratio (CSR) of the B-Mode and entropy-generated images. Entropy was measured in three situations: pre-HIFU, during- HIFU and post-HIFU (after 10 min) exposures. Considering post- and pre-HIFU exposures and entropy results, the lesion location was estimated for different acoustic powers of 30, 50, 70, 90, and 110 Watts.


  Materials and Methods Top


Image-guided high-intensity focused ultrasound system

A full description of the custom-built image-guided HIFU system used in this study has been given elsewhere.[11],[12] A 192-element confocal endocavity convex array probe (EC9-5/10, Ultrasonix) of the ultrasound imaging system (Sonix RP, Ultrasonix, Richmond, BC, Canada) operating at a center frequency of 6.5 MHz with a bandwidth of 3 MHz, was fixed in a confocal configuration in a hole in the center of the HIFU transducer to perform real-time imaging and RF echo data collecting. The HIFU transducer is a single-element concave piezo-composite transducer operating at 1 MHz with an aperture diameter of 125 mm and focal depth of 100 mm. The lateral width and axial length of the HIFU focal volume were measured in water to be 2 and 8 mm at full width at half maximum, respectively.[12] Considered size of the ROI in this study is based on the focal depth. The considered size of ROI includes the lesion and its surrounding area.

Data collection

Five fresh ex vivo porcine muscle tissue samples were exposed to HIFU ablation at different total acoustic powers of 30, 50, 70, 90, and 110 W, as described in a previous publication.[11] At each acoustic power, the RF echo data and B-Mode images were recorded in three different instances of before-, during- and 10 min after HIFU exposure by a frame rate of 16 Hz. During HIFU exposure, the RF data were recorded when the HIFU transducer was briefly turned off for 120 ms to avoid acoustic and electrical interference between the HIFU transducer and the ultrasound imaging probe. The postexposure data were recorded 10 min after the end of exposure to remove the effect of cavitation microbubbles that may be generated during the HIFU exposure in the focal region.

Each acquired RF data frame contained 192 lines with 4680 samples per line. The axial step size between two consecutive data points is about 0.0192 mm which results in the overall tissue depth of 90.1 mm in this study. The sampling rate of the RF echo data was 40 MHz and two neighboring scan lines were 0.33 mm apart.[11]

High-intensity focused ultrasound experiments

The results of this study were obtained from five ex vivo porcine muscle tissue samples. [Figure 1] shows the tissue samples after HIFU ablation when they were cut open to show the induced thermal lesions.[11]
Figure 1: Tissue samples after high-intensity focused ultrasound ablation at different acoustic powers of 30, 50, 70, 90, and 110 W

Click here to view


Radio frequency time series analysis

A time series is a sequence of data points, typically consisting of successive measurements made over a time interval. Time series analysis comprises methods for analyzing time series data to extract meaningful statistics and other characteristics of the data.[8],[13]

In this study, it is hypothesized that if a specific region of the tissue undergoes to structural changes due to HIFU energy exposure, the time series of backscattered RF echo signals from the region would carry the information pertinent to tissue structural properties. This information differs for the coagulated and noncoagulated regions of tissue.[7],[14]

In RF signal time series analysis, the values of RF echo signals measured over time from a fixed point of tissue in three instances of pre-HIFU, during-HIFU and post-HIFU (after 10 min) exposures were used to form the time series signal of each section. The length of time series signal related to each instance of data collection is equal to the corresponding frame number. [Figure 2] shows the definition of time series used in this study.
Figure 2: Samples of radio frequency echo signals, collected over time from a fixed spot of tissue under continuous emission of ultrasound, form one radio frequency signal time series

Click here to view


Shannon entropy

The present study considered entropy parameter to enhance the HIFU lesion detection in the tissue. In the information theory, the measurement of the average uncertainty of time-series in a signal is considered an entropy function.[9] To this end, the Shannon Entropy H, as given in Equation (1), is defined as a negative summation of the probability of a given random amplitude (pi) multiplied by the base 2 logarithm of (pi).[9]



In signal analysis theory, the entropy is a measure of similarity of the sample values of a signal. When all amplitude values in time series signals are the same, the value of entropy is zero and when every amplitude value is unique, the value of entropy is 1.[15]

It is well-accepted that the distribution of the ultrasound RF echo time series signals conform to the Rayleigh distribution.[16] In this research, the entropies of time series before-, during- and 10-minute after HIFU exposure were calculated in each point of the selected tissue region individually. The entropy results produced a 2D matrix with dimensions representing the ROI in tissue which is shown as a 2D image. The resulting matrix values of during-and post-exposures were divided by corresponding resulting matrix values of pre-exposure to improve the image contrast.

Contrast-to-speckle ratio

The accuracy of the HIFU lesion detectability can be quantified using the CSR parameter as proposed by Patterson et al.[17] CSR measurements are used to make a comparison between B-mode images and the results of the proposed method. The CSR is calculated by defining two equal sized regions of interest as given by equation (2):



Where S0 and are the average and variance of the signal values outside the thermal lesion, respectively. Si and are the average and variance of the signal values inside the thermal lesion at the same window size, respectively.

Nakagami distribution

In previous studies, it has been demonstrated that the Nakagami parameter and the scaling parameter of the Nakagami distribution can be effectively used to detect HIFU induced thermal lesions.[18],[19] In addition, it was reported that the scaling parameter of the Nakagami distribution led to better results in lesion detection compared with the Nakagami parameter.[19] To compare our method in estimating the location and dimension of the thermal lesion, in this study, we computed the scaling parameter of the Nakagami distribution for the first frame of the post-HIFU and pre-HIFU exposures for each acoustic power. Equation (3) gives the Nakagami probability density function.[20]





Where the symbol r denotes possible values for the random variable R of the envelope of the backscattered signal, Γ (.) is the gamma function, m is the Nakagami parameter, and Ω is the scaling parameter. m and Ω are given by the following equations:





In equations (4) and (5), E (.) is the statistical average and R represents the envelope of the backscattered signal.


  Results Top


[Figure 3] and [Figure 4] show the results of B-mode images and their corresponding entropy images in two instances of during- and post-exposure, respectively. According to [Figure 3] and [Figure 4] in the ablated region of both during- and post-exposure, decreasing the amount of entropy is seen. The lower value of entropy could be due to the changed tissue structure in effect of HIFU exposure. Furthermore, it can be said that as the tissue temperature increases, tissue particles get more dependent and becomes more homogeneous. Hence, backscattered RF echo amplitudes become more uniform in comparison to before exposure when backscattered RF echo amplitudes are more random in nature.
Figure 3:During high intensity focused ultrasound exposure images (a) B-mode images of selected region of interest at different acoustic powers of 30, 50, 70, 90, and 110 W. (b) Generated images based on the Shannon entropy

Click here to view
Figure 4: Post high intensity focused ultrasound exposure images (a) B-mode images of selected region of interest at different acoustic powers of 30, 50, 70, 90, and 110 W. (b) Generated images based on the Shannon entropy

Click here to view


To prepare data for analyzing the scaling parameter, the Hilbert transform of each scan line was taken to obtain the envelope image. The scaling parameter of the first frame of pre- and post-HIFU exposure for each acoustic power was calculated using equation (5). The scaling images were constructed using a rectangular moving window whose axial length was equal to the wavelength of the 4 MHz ultrasound beam (0.385 mm), and the lateral width spans three scan lines (3 mm × 0.33 mm). RF time series data were used in generating the entropy image. In other words, all data frames of pre-, during-, and post-HIFU instances were used to construct a unique image at each acoustic power separately. [Figure 5] shows the result of the entropy and the scaling parameter images. [Figure 5]a is the map of the average ratio of the post-exposure scaling parameter divided by the pre-exposure value, and [Figure 5]b is the map of the post-exposure entropy value divided by the pre-exposure value. As it is seen in [Figure 5], the scaling parameter did not properly detect the location of the lesion.
Figure 5: The ratio of the estimated post- to pre-values of selected region of interest and their corresponding size of lesion at acoustic power of 90 W. (a) Scaling parameter and (b) entropy parameter maps

Click here to view


[Figure 6] and [Figure 7] compare the CSR values of B-mode and entropy images at two instances of during- and post-HIFU exposures, respectively. The CSR is calculated by defining two regions of the same size inside and outside the thermal lesion in each instance. During- and post-HIFU window sizes are 51 pixels axially and 3 lines laterally equal to at each stage in the ROI. [Figure 6] and [Figure 7] demonstrate the improvement of CSR values using an entropy imaging method over the B-mode imaging during- and post-HIFU exposure. The proposed method provided better CSR in each acoustic power. In the acoustic power of 110 W, for example, the CSR value during ablation in entropy images was 10.3489, almost 3.4 times greater than B-mode images with a CSR value of 3.0536.
Figure 6: Contrast-to-speckle ratio values of B-mode images and entropy images at different acoustic powers of 30, 50, 70, 90, and 110 W

Click here to view
Figure 7: Contrast-to-speckle ratio values of B-mode images and entropy images at different acoustic powers of 30, 50, 70, 90, and 110 W

Click here to view



  Discussion Top


Results obtained in this study, demonstrate that the entropy imaging method could result in higher CSR values over the standard B-mode images and the Nakagami scaling images reported by Rangraz et al.[19] In entropy imaging, the thermal lesions were fairly well detectable both during- and post-HIFU exposure. As mentioned earlier, in equation (1) entropy is a negative summation of the probability of a signal multiplied by the base 2 logarithm of the probability. This means that unlike the standard B-mode and Nakagami imaging, the amplitude value of the signal does not have any effect on the outcome. However, in this study, to construct the B-mode and Nakagami images, the envelope of the temporal signal obtained by calculating the Hilbert transform of each scan line. In addition, the effect of different acoustic powers in tissue coagulation could be assessed in entropy generated images.

The main reason for studying the 10 min post exposure was minimizing the effects of cavitation, tissue degassing and boiling bubbles that are typically generated in the focal region during the HIFU exposure, especially at higher acoustic power levels. The induced gas bodies will scatter, reflect or distort the HIFU beam and could lead to a false determination of the exposed region position during HIFU exposure.


  Conclusion Top


In this study, a new method of RF echo signal time series analysis using the entropy parameter was presented. The main objective of this study was to detect HIFU-induced thermal lesions in an ex vivo tissue model at different total acoustic powers. Based on the results obtained, the generated entropy images provide higher contrast in detecting HIFU thermal lesions in comparison with the B-mode images at each acoustic power. Moreover, the proposed entropy method yields higher thermal lesion detectability comparing to the Nakagami imaging.

Further studies are suggested to evaluate the effect of temperature by adding a quantitative thermometry method such as thermocouples. Moreover, extending the study to in vivo animal studies which includes blood flow and tissue motion effects is proposed.

Acknowledgments

We would like to thank Arthur Worthington and Pooya Sobhe Bidari, both from Dept. of Physics, Ryerson University, Toronto, Canada, for their help in experimental set up and data acquisition. This work was partially supported by the Natural Sciences and Engineering Research Council of Canada Discovery Grant that was awarded to J. Tavakkoli.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

 
  References Top

1.
Jolesz FA, Hynynen K. MRI-Guided Focused Ultrasound Surgery. New York: Informa Healthcare; 2007.  Back to cited text no. 1
    
2.
Tavakkoli J, Sanghvi NT. Ultrasound-guided HIFU and thermal ablation. In: Frenkel V, editor. Therapeutic Ultrasound: Mechanisms to Applications. Hauppauge: Nova Science Publishers; 2011. p. 137-61.  Back to cited text no. 2
    
3.
Ebbini ES, ter Haar G. Ultrasound-guided therapeutic focused ultrasound: Current status and future directions. Int J Hyperthermia 2015;31:77-89.  Back to cited text no. 3
[PUBMED]    
4.
Damianou C, Pavlou M, Velev O, Kyriakou K, Trimikliniotis M. High intensity focused ultrasound ablation of kidney guided by MRI. Ultrasound Med Biol 2004;30:397-404.  Back to cited text no. 4
[PUBMED]    
5.
Zhou YF. High intensity focused ultrasound in clinical tumor ablation. World J Clin Oncol 2011;2:8-27.  Back to cited text no. 5
[PUBMED]    
6.
Zhong H, Wan MX, Jiang YF, Wang SP. Monitoring imaging of lesions induced by high intensity focused ultrasound based on differential ultrasonic attenuation and integrated backscatter estimation. Ultrasound Med Biol 2007;33:82-94.  Back to cited text no. 6
[PUBMED]    
7.
Moradi M, Mousavi P, Isotalo PA, Siemens DR, Sauerbrei EE, Abolmaesumi P. A New Approach to Analysis of RF Ultrasound Echo Signals for Tissue Characterization: Animal Studies. Medical Imaging 2007: Ultrasonic Imaging and Signal Processing; 2007. p. 6513.  Back to cited text no. 7
    
8.
Imani F, Abolmaesumi P, Wu MZ, Lasso A, Burdette EC, Ghoshal G, et al. Ultrasound-guided characterization of interstitial ablated tissue using RF time series: Feasibility study. IEEE Trans Biomed Eng 2013;60:1608-18.  Back to cited text no. 8
[PUBMED]    
9.
Hughes M. Analysis of digitized waveforms using Shannon entropy. J Acoust Soc Am 1993;93:892-906.  Back to cited text no. 9
    
10.
Hughes M. Analysis of ultrasonic waveforms using Shannon entropy. In: Proceedings of IEEE 1992 Ultrasonics Symposium. Vol. 20-23. Tucson, AZ, USA; 1992. p. 1205-9.  Back to cited text no. 10
    
11.
Rangraz P, Behnam H, Sobhebidari P, Tavakkoli J. Real-time monitoring of high-intensity focused ultrasound thermal therapy using the manifold learning method. Ultrasound Med Biol 2014;40:2841-50.  Back to cited text no. 11
[PUBMED]    
12.
Rahimian S, Tavakkoli J. Estimating dynamic changes of tissue attenuation coefficient during high-intensity focused ultrasound treatment. J Ther Ultrasound 2013;1:14.  Back to cited text no. 12
[PUBMED]    
13.
Imani F, Wu MZ, Lasso A, Burdette EC, Daoud M, Fitchinger G, et al. Monitoring of tissue ablation using time series of ultrasound RF data. Med Image Comput Comput Assist Interv 2011;14:379-86.  Back to cited text no. 13
[PUBMED]    
14.
Mobasheri S, Behnam H, Rangraz P, Tavakkoli J. Radio frequency ultrasound time series signal analysis to evaluate high-intensity focused ultrasound lesion formation status in tissue. J Med Signals Sens 2016;6:91-8.  Back to cited text no. 14
[PUBMED]  [Full text]  
15.
Tsui PH. Ultrasound detection of scatterer concentration by weighted entropy. Entropy 2015;17:6598-616.  Back to cited text no. 15
    
16.
Burckhardt CB. Speckle in ultrasound B-mode scans. IEEE Trans Sonics Ultrason1978;25:1-6.  Back to cited text no. 16
    
17.
Patterson MS, Foster FS. The improvement and quantitative assessment of B-mode images produced by an annular array/cone hybrid. Ultrason Imaging 1983;5:195-213.  Back to cited text no. 17
[PUBMED]    
18.
Ho MC, Lin JJ, Shu YC, Chen CN, Chang KJ, Chang CC, et al. Using ultrasound nakagami imaging to assess liver fibrosis in rats. Ultrasonics 2012;52:215-22.  Back to cited text no. 18
[PUBMED]    
19.
Rangraz P, Behnam H, Tavakkoli J. Nakagami imaging for detecting thermal lesions induced by high-intensity focused ultrasound in tissue. Proc Inst Mech Eng H 2014;228:19-26.  Back to cited text no. 19
[PUBMED]    
20.
Zhang S, Zhou F, Wan M, Wei M, Fu Q, Wang X, et al. Feasibility of using nakagami distribution in evaluating the formation of ultrasound-induced thermal lesions. J Acoust Soc Am 2012;131:4836-44.  Back to cited text no. 20
[PUBMED]    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5], [Figure 6], [Figure 7]


This article has been cited by
1 Ultrasound Entropy Imaging for Detection and Monitoring of Thermal Lesion During Microwave Ablation of Liver
Xiejing Li, Xin Jia, Ting Shen, Mengke Wang, Guang Yang, Hua Wang, Qinli Sun, Mingxi Wan, Siyuan Zhang
IEEE Journal of Biomedical and Health Informatics. 2022; 26(8): 4056
[Pubmed] | [DOI]
2 Thermometry using entropy imaging of ultrasound radio frequency signal time series
Ashkan Behnia, Hamid Behnam, Elyas Shaswary, Jahan Tavakkoli
Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 2022; : 0954411922
[Pubmed] | [DOI]
3 Recognition of Biological Tissue Denaturation Based on Improved Multiscale Permutation Entropy and GK Fuzzy Clustering
Ziqi Peng, Xian Zhang, Jing Cao, Bei Liu
Information. 2022; 13(3): 140
[Pubmed] | [DOI]
4 Recognition study of denatured biological tissues based on multi-scale rescaled range permutation entropy
Bei Liu, Wenbin Tan, Xian Zhang, Ziqi Peng, Jing Cao
Mathematical Biosciences and Engineering. 2022; 19(1): 102
[Pubmed] | [DOI]
5 In Vivo Non-Thermal, Selective Cancer Treatment With High-Frequency Medium-Intensity Focused Ultrasound
Yongkui Tang,Leng-Ying Chen,Ailin Zhang,Chun-Peng Liao,Mitchell Eric Gross,Eun Sok Kim
IEEE Access. 2021; 9: 122051
[Pubmed] | [DOI]
6 Detection of gaps between high-intensity focused ultrasound (HIFU)-induced lesions using transient axial shear strain elastograms
Chenhui Liu,Yufeng Zhou
Medical Physics. 2018;
[Pubmed] | [DOI]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
Abstract
Introduction
Materials and Me...
Results
Discussion
Conclusion